Pure Air System(Controlled Environmental Solutions)ISO-9001-2015 Certified Company
Mobile 09840506589
info@pureairindia.com
0
0

Class II Type B2 Biosafety Cabinet Manufacturers in Chennai

Class II Type B2 Cabinets:

Among Class II cabinets, Class II Type B2 cabinets provide the maximum level of protection. With a unique circulation pattern that draws air downward across the work surface and exhausts it via HEPA filters, they safeguard both employees and products. This “total exhaust” design guarantees that all of the air within the cabinet is securely filtered and evacuated. Class II Type B2 cabinets are appropriate for handling dangerous biological agents and volatile chemicals.

Working Area – Stainless Steel IS 304 Grade Satin Finish.
Product Bio safety Cabinet – Class II Type B2
Air balancing 0% recirculation.100% exhaust
Protection Product & Personnel & Environmental protection
Bio safety Level Bio safety Level 1 , 2 & 3
Exhaust Duct Requirement Hard Duct Only
Control Plenum surrounded by Negative Plenum
Cleanliness level ISO class 5 as per ISO 14644-1
Average Airflow Inflow Velocity – Minimum 0.45 m/s ( 90 FPM) at 8” front opening
Velocity Down Velocity – 0.30 m/s ( 65 FPM)
Noise level 60 db on A scale
Light Intensity Minimum 900 lux inside the working area at ambient zero
MOC Main Body – Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area – Stainless Steel IS 304 Grade Satin Finish
Supply Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
Exhaust Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
Pre Filter “Pureair” make, EU 4 rating with suitable size.
Efficiency 90% down to 10 Microns.
Noise Level Less than 70 db ‘A’.
Vibration Level Instrumentation Minimum.
Differential Pressure Gauge Magnehelic Gauge – 01 No. (make : Dwyer, USA) with each unit (To measure differential pressure across the supply HEPA filter).
Motor Blower Assembly Statically & Dynamically balanced with suitable rating and size.
Supply & Exhaust (Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise.
3 Pins Switch & Socket 15 AMPS – 1 set in each unit.
Other Switches On/Off switches for motor blower & CFL lights.
Microprocessor Control Airflow display and UV Interlocking and Low airflow Alarm and Supply & Exhaust filter Fail

Enquire for more details : info@pureairindia.com