Working Area – Stainless Steel IS 304 Grade Satin Finish
Product | Class II Biosafety Cabinet –Type A2 |
Air balancing | 70% recirculation.30% exhaust |
Protection | Product & Personnel & Environmental protection |
Bio safety Level | Bio safety Level 1 , 2 & 3 |
Exhaust Requirement | Return to Laboratory or Thimble Ducting Only |
Cleanliness level | ISO class 5 as per ISO 14644-1 |
Average Airflow | Inflow Velocity Minimum 0.45 m/s ( 90 FPM) at 8” front opening |
Velocity | Down Velocity 0.30 m/s ( 65 FPM) |
Noise level | 60 db on A scale |
Light Intensity | Minimum 900 lux inside the working area at ambient zero |
MOC | Main Body – Galvanized Iron with PU coated / Stainless Steel – Satin Finish Working Area – Stainless Steel IS 304 Grade Satin Finish |
Supply Filter | HEPA Filter having EU 13 rating with suitable size. Efficiency 99.99% down to 0.3 Microns. |
Exhaust Filter | HEPA Filter having EU 13 rating with suitable size. Efficiency 99.99% down to 0.3 Microns. |
Pre Filter | “Pureair” make, EU 4 rating with suitable size. Efficiency 90% down to 10 Microns. |
Noise Level | Less than 70 dB ‘A’. |
Vibration | Level Minimum. Instrumentation |
Magnehelic Differential Pressure Gauge | 01 No. (make: Dwyer, USA) with each unit (To measure differential pressure across the supply HEPA filter). Statically & Dynamically balanced with suitable rating and size. |
Motor Blower Assembly | (Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise. |
3 Pins Switch & Socket | 15 AMPS – 1 set in each unit. |
Other Switches | On/Off switches for motor blower & CFL lights |